Remote Control
R&S
®
FPC
288
User Manual 1178.4130.02 ─ 07
Example:
//Set the volume to 40%.
SYST:AUD:VOL 0.40
Manual operation:
See
See
19.12.10
Display Lines
...........................................................................................288
CALCulate<n>:DLINe
<LinePosition>
This command defines the position of a display line.
Prerequisites for this command
●
Turn on a display line (
Suffix:
<n>
.
irrelevant
Parameters:
<LinePosition>
<numeric value>
You can use any unit you want, the R&S
FPC then converts the
unit to the currently selected unit. If you omit a unit, the
R&S
FPC uses the currently selected unit.
*RST:
- (STATe to OFF)
Default unit: -
Example:
//Position display line
CALC:DLIN:STAT ON
CALC:DLIN –20dBm
Manual operation:
See
See
CALCulate<n>:DLINe:STATe
<State>
This command turns display lines on and off.
Suffix:
<n>
.
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
OFF | 0
Example:
//Turn on display line
CALC:DLIN:STAT ON
Manual operation:
See
See
Remote Commands of the Spectrum Application