Operating the instrument
R&S
®
FE50DTR
38
Manual 1179.3180.02 ─ 02
a) At the R&S
FSV/A on the "Global Config" tab, set "Connection State" to
"Off".
The connection is aborted. The R&S
FSV/A releases the lock of the
R&S
FE50DTR.
b) At the R&S
SMW, change the frontend settings according to your setup.
How to:
"To configure frequency settings at the vector signal analyzer"
c) If you want to measure with the changed configuration of the
R&S
FE50DTR, reestablish the control connection between R&S
FSV/A
and R&S
FE50DTR ("Connection State" > "On").
To configure frequency settings
Simultaneous mode operation of the R&S
FE50DTR requires an identical setting
of the frequency band configuration at the R&S
SMW and at the R&S
FSV/A.
How to:
●
Configure frequency settings at the R&S SMW
.
●
Configure frequency settings at the R&S FSV/A
.
The frequency band configuration depends on the equipment of the vector signal
generator and vector signal analyzer.
illustrates exemplary configura-
tions of R&S
SMW and R&S
FSV/A and availability of the frequency band config-
uration modes "IF High" and "IF Low".
Table 6-4: Analyzer/signal generator equipment and frequency band configuration
Analyzer model
Analyzer
hardware
Analysis
bandwidth
Signal generator
hardware
Band configuration
R&S
FSV(A)3007
</= B400
</= 400
MHz
B1006, B1007
"IF Low"
R&S
FSV(A)3007
</= B400
</= 400
MHz
>/= B1012
"IF Low"
R&S
FSV(A)3013
R&S
FSV(A)3030
R&S
FSV(A)3044
</= B400
</= 400
MHz
B1006, B1007
"IF Low"
Operating the R&S
FE50DTR in simultaneous mode