Remote Control
R&S
®
ESRP
840
User Manual 1175.7445.02 ─ 08
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK2 ON
Switches on marker 2.
CALC:MARK:FUNC:NOIS ON
Switches on noise measurement.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK2:NOIS:RES?
Outputs the noise result of marker 2.
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:NOISe[:STATe]
<State>
This command turns the noise measurement for all markers on or off.
If on, the R&S
ESRP measures the noise power density at the marker position.
The result can be queried with
CALCulate<n>:MARKer<m>:FUNCtion:NOISe:
.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:MARK:FUNC:NOIS ON
Switches on the noise measurement.
Manual operation:
See
Phase Noise
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:AUTO
............................................... 840
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:RESult?
............................................ 841
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise[:STATe]
............................................. 841
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:AUTO
<State>
This command turns an automatic peak search for the fixed reference marker at the
end of a sweep on and off.
Suffix:
<n>
.
Selects the measurement window.
<m>
irrelevant
Remote Commands in Spectrum Analyzer Mode