Remote Control
R&S
®
ESRP
841
User Manual 1175.7445.02 ─ 08
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:DELT:FUNC:PNO:AUTO ON
Activates an automatic peak search for the reference marker in
a phase-noise measurement.
Manual operation:
See
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:RESult?
This command queries the result of the phase noise measurement.
If necessary, the command activates the measurement first..
Suffix:
<n>
.
Selects the measurement window.
<m>
2
Marker 2 is always the deltamarker for phase noise measure-
ments.
Example:
CALC:DELT2:FUNC:PNO:RES?
Outputs the result of phase-noise measurement of the delta-
marker 2.
Usage:
Query only
Manual operation:
See
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise[:STATe]
<State>
This command turns the phase noise measurement at the delta marker position on and
off.
The correction values for the bandwidth and the log amplifier are taken into account in
the measurement.
The reference marker for phase noise measurements is either a normal marker or a
fixed reference. If necessary, the command turns on the reference marker
A fixed reference point can be modified with the
CALCulate<n>:DELTamarker<m>:FUNCtion:
commands independent of the position of marker 1 and of a trace.
Suffix:
<n>
.
Selects the measurement window.
<m>
irrelevant
Note
: marker 2 is always the deltamarker for phase noise mea-
surement results.
Remote Commands in Spectrum Analyzer Mode