Remote Control
R&S
®
ESRP
818
User Manual 1175.7445.02 ─ 08
Parameters:
<State>
ON | OFF
*RST:
ON for TRACe1, OFF for TRACe2 to 6
Example:
DISP:TRAC3 ON
Manual operation:
See
FORMat[:DATA]
<Format>
This command selects the data format for the data transmitted from the R&S
ESRP to
the controlling computer. It is used for the transmission of trace data. The data format
of trace data received by the instrument is automatically recognized, regardless of the
format which is programmed.
In the "Spectrum" mode, the format setting REAL, 32 is used for the binary transmis-
sion of trace data.
Parameters:
<Format>
ASCii
ASCII data are transmitted in plain text, separated by commas.
REAL
REAL data are transmitted as 32-bit IEEE 754 floating-point
numbers in the "definite length block format".
*RST:
ASCII
Example:
FORM REAL,32
FORM ASC
FORMat:DEXPort:DSEParator
<Separator>
This command defines which decimal separator (decimal point or comma) is to be
used for outputting measurement data to the file in ASCII format. Different languages
of evaluation programs (e.g. MS-Excel) can thus be supported.
Parameters:
<Separator>
POINt | COMMA
*RST:
(factory setting is POINt; *RST does not affect set-
ting)
Example:
FORM:DEXP:DSEP POIN
Sets the decimal point as separator.
Manual operation:
See
See
See
See
Remote Commands in Spectrum Analyzer Mode