Spectrum Measurements
R&S
®
ESRP
240
User Manual 1175.7445.02 ─ 08
List Evaluation (On/Off) ← List Evaluation ← Spectrum Emission Mask
Activates or deactivates the list evaluation.
Remote command:
Turning list evaluation on and off:
CALCulate<n>:PEAKsearch|PSEarch:AUTO
Querying list evaluation results:
Margin ← List Evaluation ← Spectrum Emission Mask
Opens an edit dialog box to enter the margin used for the limit check/peak search.
Remote command:
CALCulate<n>:PEAKsearch|PSEarch:MARGin
Show Peaks ← List Evaluation ← Spectrum Emission Mask
In the diagram, marks all peaks with blue squares that have been listed during an
active list evaluation.
Remote command:
CALCulate<n>:ESPectrum:PSEarch|:PEAKsearch:PSHow
Save Evaluation List ← List Evaluation ← Spectrum Emission Mask
Opens the "ASCII File Export Name" dialog box to save the result in ASCII format to a
specified file and directory. For further details refer also to the "ASCII File Export" soft-
key (
Remote command:
ASCII File Export ← Save Evaluation List ← List Evaluation ← Spectrum Emis-
sion Mask
Opens the "ASCII File Export Name" dialog box and saves the active peak list in ASCII
format to the specified file and directory.
The file consists of the header containing important scaling parameters and a data sec-
tion containing the marker data. For details on an ASCII file see
This format can be processed by spreadsheet calculation programs, e.g. MS-Excel. It
is necessary to define ';' as a separator for the data import. Different language versions
of evaluation programs may require a different handling of the decimal point. It is there-
fore possible to select between separators '.' (decimal point) and ',' (comma) using the
"Decim Sep" softkey (see
An example of an output file for Spectrum Emission Mask measurements is given in
"ASCII File Export Format (Spectrum Emission Mask)"
Remote command:
Measurements