Remote Control
R&S
®
ESRP
672
User Manual 1175.7445.02 ─ 08
Example:
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
CALC:DELT2 ON
Switches on delta marker 2.
CALC:DELT2:Y?
Outputs measurement value of delta marker 2.
Manual operation:
See
"Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
11.4.8
Limit Lines
...............................................................................................672
......................................................................................... 673
11.4.8.1
Using Display Lines
................................................................................................672
..................................................................................... 672
CALCulate<n>:DLINe<k>
<Position>
This command defines the (vertical) position of a display line.
Suffix:
<n>
.
Selects the measurement window.
<k>
1 | 2
Selects the display line.
Parameters:
<Position>
The value range is variable.
The unit depends on
*RST:
(state is OFF)
Example:
CALC:DLIN -20dBm
Manual operation:
See
"Display Line 1 / Display Line 2"
CALCulate<n>:DLINe<k>:STATe
<State>
This command turns a display line on and off.
Suffix:
<n>
.
Selects the measurement window.
<k>
1 | 2
Selects the display line.
Remote Commands in Receiver Mode