Remote Control
R&S
®
ESRP
651
User Manual 1175.7445.02 ─ 08
Manual operation:
See
MMEMory:STORe:FINal
<FileName>
This command exports the contents of the final measurement peak list to a file in ASCII
format.
Parameters:
<FileName>
String containg the file name. The extension of the file is *.dat.
Example:
:MMEM:STOR:FIN 'A:\TEST.DAT'
Usage:
Event
Manual operation:
See
MMEMory:STORe:PEAKlist
<FileName>
This command exports the contents of a peak list to a file in ASCII format.
Parameters:
<FileName>
String containg the file name. The extension of the file is *.dat.
Example:
MMEM:STOR:PEAK 'D:\TEST.DAT'
Manual operation:
See
11.4.6.5
Trace Configuration
Commands useful to configure traces and final measurements described elsewhere:
●
●
●
[SENSe:]DETector<t>:FMEasurement
[SENSe:][WINDow:]DETector<trace>[:FUNCtion]
.............................................................. 651
DISPlay[:WINDow<n>]:TRACe<t>:MODE
.........................................................................652
[SENSe:][WINDow:]DETector<trace>[:FUNCtion]
<Detector>
This command selects the detector for the scan.
Suffix:
<trace>
.
1...6
Selects the trace.
Parameters:
<Detector>
NEGative | POSitive | RMS | AVERage | QPEak | CAVerage |
CRMS
*RST:
APEak
Example:
DET RMS
Selects the RMS detector.
Remote Commands in Receiver Mode