CONFigure:TDSCdma:MEAS<i>:MEValuation:NFRames
<NumOfFrame>
Selects the total number of measured subframes. One subframe contains 7 timeslots.
Parameters:
<NumOfFrame>
Range:
1 subframe to 16 subframes
*RST:
1 subframe
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V1.0.2.7
Manual operation:
See
CONFigure:TDSCdma:MEAS<i>:MEValuation:PSLot
<PreSelSlot>
Selects the so-called "Preselected Slot", i.e. an uplink slot number for display of single
slot measurement results, e.g. ACLR measurement, EVM vs.chip.
The allowed range of values depends on the number of measured subframes (
) and on the position of the
switching point (
select any measured uplink slot, so any slot within the measured subframes, between
slot 1 and the switching point position of the subframe.
Example: 4 measured subframes, switching point position after slot 3, so slot 1 to 3 of
each subframe can be selected, allowed value range = 1,2,3,8,9,10,15,16,17,22,23,24
See also
chapter 3.1.3, "Defining the Scope of the Measurement"
Parameters:
<PreSelSlot>
Range:
1 to 7 * <measured subframes> - 1 (only UL slots)
*RST:
1
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V1.0.2.7
Manual operation:
See
CONFigure:TDSCdma:MEAS<i>:MEValuation:SSLot
<ScalarSlot>
Selects the so-called "Scalar Slot", i.e. the number of the slot where the R&S
CMW cal-
culates "single value" results (also called "scalar results"). The allowed range of values
depends on the number of measured subframes (
).
See also
chapter 3.1.3, "Defining the Scope of the Measurement"
Multi Evaluation Measurement Commands