GUI Reference
R&S
®
CMW-KM750/-KM751
40
User Manual 1173.9670.02 ─ 05
Fig. 4-3: Measurement control settings
Repetition
Defines how often the measurement is repeated if it is not stopped explicitly or by a failed
limit check.
●
Continuous:
The measurement is continued until it is explicitly terminated; the
results are periodically updated.
●
Single-Shot:
The measurement is stopped after one statistics cycle.
Single-shot is preferable if only a single measurement result is required under fixed con-
ditions, which is typical for remote-controlled measurements. Continuous mode is suita-
ble for monitoring the evolution of the measurement results in time and observe how they
depend on the measurement configuration, which is typically done in manual control. The
reset/preset values therefore differ from each other.
Remote command:
Stop Condition
Specifies the conditions for an early termination of the measurement:
●
None:
The measurement is performed according to its "Repetition" mode and "Sta-
tistic Count", irrespective of the limit check results.
●
On Limit Failure:
The measurement is stopped as soon as one of the limits is
exceeded, irrespective of the repetition mode set. If no limit failure occurs, it is per-
formed according to its "Repetition" mode and "Statistic Count". Use this setting for
measurements that are essentially intended for checking limits, e.g. production tests.
Remote command:
Parameters and Settings