Command Reference
R&S
®
CMW-KM750/-KM751
82
User Manual 1173.9670.02 ─ 05
When the measurement has completed the first measurement cycle (first single shot),
the statistical depth is reached and the timer is reset.
If the first measurement cycle has not been completed when the timer expires, the mea-
surement is stopped. The measurement state changes to
RDY
and the reliability indicator
is set to 1, indicating that a measurement timeout occurred. Still running
READ
,
FETCh
or
CALCulate
commands are completed, returning the available results. At least for
some results there are no values at all or the statistical depth has not been reached.
A timeout of 0 s corresponds to an infinite measurement timeout.
Parameters:
<TCDTimeout>
*RST:
0 s
Default unit: s
Firmware/Software:
V2.0.10
V3.0.20: *RST added
CONFigure:TDSCdma:MEAS<i>:MEValuation:REPetition
<MeasEvlRept>
Specifies the repetition mode of the measurement. The repetition mode specifies whether
the measurement is stopped after a single-shot or repeated continuously. Use
CONFigure:..:MEAS<i>:...:SCOunt
to determine the number of measurement
intervals per single shot.
See also: "Statistical Settings" in the R&S CMW user manual, chapter "Remote Control"
Parameters:
<MeasEvlRept>
SINGleshot | CONTinuous
SINGleshot
: Single-shot measurement
CONTinuous
: Continuous measurement
*RST:
SING
Example:
See
Single-Shot and Continuous Measurements
Firmware/Software:
V1.0.2.7
Manual operation:
See
CONFigure:TDSCdma:MEAS<i>:MEValuation:SCONdition
<StopCond>
Qualifies whether the measurement is stopped after a failed limit check or continued.
OLFail
means that the measurement is stopped (
STOP: ... MEAS<i>...
) and rea-
ches the
RDY
state as soon as one of the results exceeds the limits.
Parameters:
<StopCond>
NONE | OLFail
NONE
: Continue measurement irrespective of the limit check
OLFail
: Stop measurement on limit failure
*RST:
NONE
Example:
See
Specifying Additional Measurement-Specific Settings
Multi Evaluation Measurement Commands