Command Reference
R&S
®
CMW-B400B/-B405A
92
User Manual 1176.8116.02 ─ 04
CONFigure:AUDio:SPEech<i>:ANALog:OLEVel
<OutputLevel>
Configures the D/A conversion for analog speech scenarios with an output signal at the
AF connector.
Parameters:
<OutputLevel>
Analog output signal peak voltage corresponding to digital signal
full-scale value
Range:
0 V to 7 V
*RST:
0.3 V
Default unit: V
Firmware/Software:
V3.2.10
Manual operation:
See
"Output Level Full-Scale (Peak)"
CONFigure:AUDio:SPEech<i>:CDRift
<ClockDrift>
Adjusts the clock of the audio board.
Parameters:
<ClockDrift>
Range:
-100 ppm to 100 ppm
*RST:
0 ppm
Default unit: ppm
Additional parameters: OFF | ON (disables | enables the clock
adjustment)
Firmware/Software:
V3.5.10
Manual operation:
See
CONFigure:AUDio:SPEech<i>:ANALog:FILTer:HPASs
<Frequency>
Configures the cutoff frequency of the highpass filter, located between the AF IN con-
nectors and the speech encoder.
Parameters:
<Frequency>
OFF | H6 | H50
OFF:
highpass filter disabled
H6 | H50:
cutoff frequency 6 Hz or 50 Hz
*RST:
H6
Firmware/Software:
V3.2.30
Manual operation:
See