Rohde & Schwarz CMU-Z6 Operating Instructions Manual Download Page 16

Operating Instructions

CMU-Z6

1150.0199.12

6

E-3

0

1

2

3

4

5

6

7

8

9

.

-

ON / OFF

ENTER

1100.0008. 02

VA R IA TION

D ATA

FU NC TION

SY STEM

C ONTR OL

DATA

CTRL

ME NU

SE LECT

HE LP

SETUP

m ark

sym b

E

F

B

M/ µ
µV
W

C

dBµV

k/m

G/n

A

mV

D

1

dB

dBm

UNIT.. .

EXP/ CMP

CONT/HALT

RF4 IN

UN IVERSA L  R ADI O C OMMUN ICA TION   TEST ER              CM U 200

.

ESCAPE

*

abc

def

ghi

jkl

m no

pqr

stu

vwx

yz

_ µ 

AF IN

AF OUT

VOL

DE L

AUTO

INFO

RESET

PRINT

*

#

ΩΩ

AUX 1

AUX 2

SPEECH

AUX 3

DATA 2

13 d Bm MAX

2 W MAX

50 W MAX

13 d Bm MAX

RF 3 OUT

RF 2

RF 1

DATA 1

INS

CLR

CMU200

SMIQ

REF IN

REF

RF 50 

RF 2

To check the accuracy of the corrected modulation measurement,

 Perform 

Preset

 of the SMIQ settings and a 

Reset

 of the CMU.

 

Set the SMIQ

 to comply with the CMU default analyzer settings:

FREQUENCY 1922.4 MHz

LEVEL 0 dBm

DIGITAL MOD

–> STATE

–> ON

–> SELECT STANDARD...

–> WCDMA QPSK

–> SYMBOL RATE

–> 3.84 Mcps

–> Filter

–> FILTER MODE

–> LOW_EVM

 

Set the 

input path of the 

CMU 

and start the modulation measurement:

Group Config.

–> Input Level

–> Max. Level

–> Mode

–>

Manual

–> Manual Level

–>

+5 dBm

–> Attenuation

–>

Low Noise

Connect. Control

–> Synch.

–> Reference Frequency

–> Ext. (at REF IN) 10 MHz

WCDMA 1900 FDD –> Modulation

–> Application

–> Overview QPSK

The RMS Error Vector Magnitude is shown in the second row of the 

Overview QPSK

 output table. The

result should be approx. 1.2 %:

Please bear in mind that the displayed EVM result also contains a possible inaccuracy of the generator
signal.

Summary of Contents for CMU-Z6

Page 1: ...1150 0199 12 03 1 Test and Measurement Division Operating Instructions Improvement of Wideband Modulation Analysis CMU Z6 1150 0001 02 Printed in the Federal Republic of Germany...

Page 2: ......

Page 3: ...ory environments or in the field and must not be used in any way that may cause personal injury or property damage You are responsible if the product is used for an intention other than its designated...

Page 4: ...r unassociated documentation can result in misinterpretations and thus contribute to personal injury or material damage Basic safety instructions 1 The product may be operated only under the operating...

Page 5: ...ccordingly 9 In the case of products of safety class I with movable power cord and connector operation is permitted only on sockets with earthing contact and protective earth connection 10 Intentional...

Page 6: ...roperly replaced this can cause an explosion warning lithium cells Replace the battery or storage battery only with the matching Rohde Schwarz type see spare parts list Batteries and storage batteries...

Page 7: ...a persona cosa pueda ser da ada El uso del producto fuera de sus fines definidos o despreciando las informaciones de seguridad del fabricante queda en la responsabilidad del usuario El fabricante no s...

Page 8: ...definici n Por eso se debera tener en cuenta que las palabras de se al aqu descritas sean utilizadas siempre solamente en combinaci n con la correspondiente documentaci n y solamente en combinaci n co...

Page 9: ...da En vista a la protecci n de la vida en desarrollo deber an ser protegidas personas embarazadas debidamente Tambi n las personas con un bypass pueden correr peligro a causa de la radiaci n electroma...

Page 10: ...y la conexi n deber n ser efecutadas por un electricista especializado 20 En caso de que los productos que son instalados fijamente en un lugar sean sin protector implementado autointerruptor o simila...

Page 11: ...ejemplo paredes y estantes 31 Las asas instaladas en los productos sirven solamente de ayuda para el manejo que solamente est previsto para personas Por eso no est permitido utilizar las asas para la...

Page 12: ......

Page 13: ...mode REF IN RF OUT DC Supply Rohde Schwarz CMU Z6 1150 0001 02 0 1 A 1 0 B 1 1 C MODE SETTING ON Input for 110 8 MHz reference frequency DC power supply Output for test signal LEDs indicating the ope...

Page 14: ...with the 5 V DC power from the AUX connector at the rear panel of the CMU Use the two N to BNC cables to provide CMU Z6 with a 110 8 MHz reference frequency from RF3 OUT and to feed the generated test...

Page 15: ...ormal operating temperature of the CMU otherwise the correction values might be impaired Checking the Corrected Measurement The accuracy of the corrected modulation measurement can be checked by using...

Page 16: ...on measurement Perform a Preset of the SMIQ settings and a Reset of the CMU Set the SMIQ to comply with the CMU default analyzer settings FREQUENCY 1922 4 MHz LEVEL 0 dBm DIGITAL MOD STATE ON SELECT S...

Reviews: