9.  Derating Curve 

The  derating  curve  of  the  absolute  maximum  input  voltage  in  common  mode  is  shown  as 

follows and valid for both models RT-ZD002 and RT-ZD003. 

 

 
 

10.  Inspection Procedure

 

a.  Connect  the  BNC  output  connector  to  the  vertical  input  of  a  general  purpose 

oscilloscope. 

b.  Install four AA cells or connect an appropriate power lead. 

c.  Set the oscilloscope input coupling to 1 M

, DC and to 1V/div. Center the trace on the 

display. 

d.  Connect the inputs of the probe to power lines, using an appropriate power socket to   

4 mm adapter. 

e.  Set the range of the probe to 1/100 (RT-ZD002) or 1/200 (RT-ZD003). 

f.  Then, a 50 Hz/ 60 Hz sine-wave of proper amplitude will be displayed on the screen of 

the oscilloscope and this means the probe is working properly. 

11.  Cleaning 

Use a soft cloth to clean the dirt. Prevent damage to probe. 

a.  Avoid immersing the probe. 

b.  Avoid using abrasive cleaners. 

c.  Avoid using chemicals contains benzene or similar solvents. 

 

 

1000

1000

700

150

100

1000

10000

1,0E+06

1,0E+07

1,0E+08

V

o

lt

ag

(V

o

lt

 r

m

s)

Frequency (Hz)

V

o

lt

ag

(V

 r

m

s)

 

Summary of Contents for 1337.9700.02

Page 1: ...INSTRUCTION MANUAL RT ZD002 RT ZD003 25 MHz Active Differential Probes RT ZD002 1337 9700 02 RT ZD003 1337 9800 02...

Page 2: ...injury and prevent damage to this probe or any products connected to it Observe Maximum Working Voltage To avoid any injury do not use the probe under the condition that the voltage between both input...

Page 3: ...ard do not operate this probe in an explosive atmosphere Avoid Exposed Circuit To avoid injury remove jewelry such as rings watches and other metallic objects Do not touch exposed connections and comp...

Page 4: ...00 V or 1400 V WARNING To protect against electric shock use only the accessories supplied with this probe d Using the appropriate probe accessories connect the inputs to the circuit under test CAUTIO...

Page 5: ...maged or can t be operated properly Polarity is inside and outside For wrong polarity a built in circuit protects the probe no danger or damage will occur When the voltage of the cells become too low...

Page 6: ...C or 1000 Vrms Maximum Input Voltage each side to ground 1000 V CAT III 1000 V CAT III Output Voltage Swing 7 V into 50 k load 7 V into 50 k load Output Voltage Offset 5 mV meas 5 mV meas Output Volta...

Page 7: ...splay d Connect the inputs of the probe to power lines using an appropriate power socket to 4 mm adapter e Set the range of the probe to 1 100 RT ZD002 or 1 200 RT ZD003 f Then a 50 Hz 60 Hz sine wave...

Page 8: ...de and Schwarz please visit our homepage on the R S Internet http www rohde schwarz com For queries regarding technical aspects of our products please contact our customer support at http www customer...

Page 9: ...Messtechnik GmbH Kelsterbacher Strasse 15 19 60528 Frankfurt am Main Tel 49 0 69 67724 583 Fax 49 0 69 67724 582 info allice de www allice de Verwendete Warenzeichen und Schutzrechte sind Eigentum der...

Reviews: