8. Specifications
RT-ZD002
RT-ZD003
Bandwidth
DC to 25 MHz (-3 dB)
DC to 25 MHz (-3 dB)
Attenuation Ratio
1:10 / 1:100
1:20 / 1:200
Accuracy
±2 % (meas.)
±2 % (meas.)
Rise Time
14 ns (calculated from bandwidth)
14 ns (calculated from bandwidth)
Input Impedance
8 M
Ω
// 2.75 pF differential
4 M
Ω
// 5.5 pF each side to ground
8 M
Ω
// 2.75 pF differential
4 M
Ω
// 5.5 pF each side to ground
Input Voltage
- Differential Range
1:10 ±70 V (DC+peak AC)
1:100 ±700 V (DC+peak AC)
1:20 ±140 V (DC+peak AC)
1:200 ±1400 V (DC+peak AC)
or 1000 Vrms
Input Voltage
- Common Mode Range
1:10 and 1:100
±700 V (DC+peak AC)
1:20 and 1:200
±1400 V (DC+peak AC)
or 1000 Vrms
Maximum Input Voltage
(each side to ground)
1000 V CAT III
1000 V CAT III
Output Voltage Swing
±7 V (into 50 k
Ω
load)
±7 V (into 50 k
Ω
load)
Output Voltage Offset
±5 mV (meas.)
±5 mV (meas.)
Output Voltage Noise
0.7 mVrms (meas.)
0.7 mVrms (meas.)
Oscilloscope Input
Impedance
1 M
Ω
1 M
Ω
CMRR (meas.)
-86 dB @ 50 Hz, -66 dB @ 20 kHz
-80 dB @ 50 Hz, -60 dB @ 20 kHz
Operating Temperature -10
°
C to 40
°
C
-10
°
C to 40
°
C
Storage Temperature -30 °C to 70
°
C
-30 °C to 70
°
C
Humidity
25 % to 85 % RH
25 % to 85 % RH
Power Requirements
- Standard
4 x AA Cells
4 x AA Cells
Power Requirements
- Optional
4.5 V to 12 V, 360 mW
4.5 V to 12 V, 360 mW
Length of BNC Cable
95 cm
95 cm
Length of Input Leads
45 cm
45 cm
Weight
400 g (probe and PVC jacket)
400 g (probe and PVC jacket)
Dimensions (LxWxH)
170 mm x 63 mm x 21 mm
170 mm x 63 mm x 21 mm