8. Specifications 

 

 

RT-ZD002 

RT-ZD003 

Bandwidth 

DC to 25 MHz (-3 dB) 

DC to 25 MHz (-3 dB) 

Attenuation Ratio 

1:10 / 1:100 

1:20 / 1:200 

Accuracy 

±2 % (meas.) 

±2 % (meas.) 

Rise Time 

14 ns (calculated from bandwidth) 

14 ns (calculated from bandwidth) 

Input Impedance 

8 M

 // 2.75 pF differential 

4 M

 //    5.5 pF each side to ground 

8 M

 // 2.75 pF differential 

4 M

 //    5.5 pF each side to ground 

Input Voltage 
- Differential Range 

1:10 ±70 V (DC+peak AC)   
1:100 ±700 V (DC+peak AC)   
 

1:20 ±140 V (DC+peak AC)   
1:200 ±1400 V (DC+peak AC)   
or 1000 Vrms 

Input Voltage 
- Common Mode Range

 

1:10 and 1:100   
±700 V (DC+peak AC)

 

1:20 and 1:200   
±1400 V (DC+peak AC)   
or 1000 Vrms

 

Maximum Input Voltage 
(each side to ground) 

1000 V CAT III 

1000 V CAT III 

Output Voltage Swing

 

±7 V (into 50 k

 

load) 

±7 V (into 50 k

 

load)

 

Output Voltage Offset

 

±5 mV (meas.)

 

±5 mV (meas.)

 

Output Voltage Noise 

0.7 mVrms (meas.) 

0.7 mVrms (meas.)

 

Oscilloscope Input 
Impedance 

1 M

 

1 M

 

CMRR (meas.) 

-86 dB @ 50 Hz, -66 dB @ 20 kHz 

-80 dB @ 50 Hz, -60 dB @ 20 kHz 

Operating Temperature         -10 

°

C to 40 

°

-10 

°

C to 40 

°

Storage Temperature            -30 °C to 70 

°

-30 °C to 70 

°

Humidity                             

25 % to 85 % RH 

25 % to 85 % RH 

Power Requirements 
- Standard 

4 x AA Cells 

4 x AA Cells 

Power Requirements 
- Optional 

4.5 V to 12 V, 360 mW 

4.5 V to 12 V, 360 mW 

Length of BNC Cable                                                                 

95 cm 

95 cm 

Length of Input Leads                             

45 cm 

45 cm 

Weight 

400 g (probe and PVC jacket)   

400 g (probe and PVC jacket)   

Dimensions (LxWxH)         

170 mm x 63 mm x 21 mm 

170 mm x 63 mm x 21 mm 

 

 

 

Summary of Contents for 1337.9700.02

Page 1: ...INSTRUCTION MANUAL RT ZD002 RT ZD003 25 MHz Active Differential Probes RT ZD002 1337 9700 02 RT ZD003 1337 9800 02...

Page 2: ...injury and prevent damage to this probe or any products connected to it Observe Maximum Working Voltage To avoid any injury do not use the probe under the condition that the voltage between both input...

Page 3: ...ard do not operate this probe in an explosive atmosphere Avoid Exposed Circuit To avoid injury remove jewelry such as rings watches and other metallic objects Do not touch exposed connections and comp...

Page 4: ...00 V or 1400 V WARNING To protect against electric shock use only the accessories supplied with this probe d Using the appropriate probe accessories connect the inputs to the circuit under test CAUTIO...

Page 5: ...maged or can t be operated properly Polarity is inside and outside For wrong polarity a built in circuit protects the probe no danger or damage will occur When the voltage of the cells become too low...

Page 6: ...C or 1000 Vrms Maximum Input Voltage each side to ground 1000 V CAT III 1000 V CAT III Output Voltage Swing 7 V into 50 k load 7 V into 50 k load Output Voltage Offset 5 mV meas 5 mV meas Output Volta...

Page 7: ...splay d Connect the inputs of the probe to power lines using an appropriate power socket to 4 mm adapter e Set the range of the probe to 1 100 RT ZD002 or 1 200 RT ZD003 f Then a 50 Hz 60 Hz sine wave...

Page 8: ...de and Schwarz please visit our homepage on the R S Internet http www rohde schwarz com For queries regarding technical aspects of our products please contact our customer support at http www customer...

Page 9: ...Messtechnik GmbH Kelsterbacher Strasse 15 19 60528 Frankfurt am Main Tel 49 0 69 67724 583 Fax 49 0 69 67724 582 info allice de www allice de Verwendete Warenzeichen und Schutzrechte sind Eigentum der...

Reviews: