![Rohde & Schwarz 1330.5000K04 Getting Started Download Page 85](http://html1.mh-extra.com/html/rohde-and-schwarz/1330-5000k04/1330-5000k04_getting-started_1477780085.webp)
Operating the Instrument
R&S
®
FSVA3000/R&S
®
FSV3000
85
Getting Started 1330.8073.02 ─ 07
1
3
4
5
2
(1)
Trace color
Color of trace display in diagram
(2)
Trace no.
Trace number (1 to 6)
(3)
Detector
Selected detector:
AP
AUTOPEAK detector
Pk
MAX PEAK detector
Mi
MIN PEAK detector
Sa
SAMPLE detector
Av
AVERAGE detector
Rm
RMS detector
QP
QUASIPEAK detector
(4)
Trace Mode
Sweep mode:
Clrw
CLEAR/WRITE
Max
MAX HOLD
Min
MIN HOLD
Avg
AVERAGE (Lin/Log/Pwr)
View
VIEW
(5)
Smoothing factor
Smth
Smoothing factor, if enabled.
Norm/NCor
Correction data is not used.
7.1.3
Marker Information
Marker information is provided either in the diagram grid or in a separate marker
table, depending on the configuration.
Marker information in diagram grid
Within the diagram grid, the x-axis and y-axis positions of the last two markers or
delta markers that were set are displayed, if available, as well as their index. The