![Rohde & Schwarz 1330.5000K04 Getting Started Download Page 83](http://html1.mh-extra.com/html/rohde-and-schwarz/1330-5000k04/1330-5000k04_getting-started_1477780083.webp)
Operating the Instrument
R&S
®
FSVA3000/R&S
®
FSV3000
83
Getting Started 1330.8073.02 ─ 07
cable for the current measurement, as opposed to the channel-specific settings
that are always displayed.
The following types of information are displayed, if applicable.
Table 7-2: Common settings displayed in the channel bar
"SGL"
The sweep is set to single sweep mode.
"Sweep
Count"
The current signal count for measurement tasks that involve a specific number
of subsequent sweeps
(see "Sweep Count" setting in "Sweep settings" in the User Manual)
"TRG"
Trigger source
(for details see "Trigger settings" in the User Manual)
●
EXT
: External
●
IFP
: IF power (+trigger bandwidth)
●
PSE
: Power sensor
●
RFP
: RF power
●
SQL
: Squelch
●
TIM
: Time
●
VID
: Video
"6dB"/"RRC"
/"CHN"
Filter type for sweep bandwidth
(see "Bandwidth settings" in the User Manual)
"PA"
/
Ext
"PA"
The preamplifier is activated. / Data compensation is performed using data
from the (optional) external preamplifier.
"YIG
Bypass"
The YIG filter is deactivated.
"GAT"
The frequency sweep is controlled via the TRIGGER INPUT connector.
"TDF"
The specified transducer factor is activated.
"75
Ω
"
The input impedance of the instrument is set to 75
Ω
.
"FRQ"
A frequency offset
≠
0 Hz is set.
"DC/AC"
DC or AC coupling is used for the input.
"ExtMix"
<band>
An external mixer is activated for input (requires option R&S
FSV3-B21); the
used band is also indicated
"Ext. Gen"
"Ext. Gen":
The R&S
FSV/A is controlling an external generator (requires
option R&S
FSV3-B10).