![Rohde & Schwarz 1164.4391.03 Service Manual Download Page 139](http://html1.mh-extra.com/html/rohde-and-schwarz/1164-4391-03/1164-4391-03_service-manual_1477763139.webp)
R&S FSP
Module Replacement
1164.4556.82 3.35
E-1
Installing the New Modules and Completing the Instrument
Mount the new Key-Probe board (50) with three countersunk screws (55) to mounting plate (60.)
Mount the new Vol./Phone Board (1040) with the nut of the volume control and one countersunk
screw (1055) to the mounting plate (1050). Mount Vol./Phone board (1040) with three countersunk
screws (1060) to mounting plate (60).
Connect the ribbon cable at motherboard-connector X80 and X81.
Move the complete mounting plate (60) carefully back in the instrument.
CAUTION
Make sure to route the cables properly.
Connect RF-cable W1 (295), (315) or (322) at the input of the RF Attenuator (20) or (30).
Screw three countersunk screws (70) to the instrument frame and two countersunk screws (70) to
the mounting plate (60).
Install the labelling panel (270).
Push the 2 front handles (420) on the instrument and mount them with the 4 screws (430).
Replace the volume control knob (Option R&S FSP-B3), if it was removed.
Push the tube (410) on the instrument and mount the 4 rear-panel feet (450).
Function Test
Connect a suitable keyboard to the keyboard jack.
Connect the instrument to the mains, switch on the instrument power switch and switch on the
instrument with the ON/STANDBY key
Press the "Ctrl" and "Esc" keys on the keyboard simultaneously; the Windows-XP taskbar appears at
the bottom of the screen.
Measure the output voltage at the probe connector (see Fig. 3-12
Pin assignments of PROBE
POWER connector).
Connect a keyboard to the instrument and test the function.
If the option R&S FSP-B3 (AF demodulator) is fitted, test the function of headphone connector and
volume control. Press the
MKR
key, and
press the
MARKER DEMOD
softkey
.
Noise can be heard in
the loudspeaker, the volume can be changed with the
VOLUME
knob at the front panel. Connect a
headphone to the
AF OUTPUT
at the front panel, the loudspeaker must be switched off and the
noise can be heard in the headphone.
Pin Signal
1
GND
2
-12.6 V
1
2
3
3
+15 V
Fig. 3-12
Pin assignments of PROBE POWER connector
Summary of Contents for 1164.4391.03
Page 4: ......
Page 14: ......
Page 16: ......
Page 18: ......
Page 20: ......
Page 22: ......
Page 24: ......
Page 92: ...Contents Adjustment R S FSP 1164 4556 82 I 2 2 E 1 This page is left blank intentionally...
Page 103: ...R S FSP Contents Repair 1164 4556 82 I 3 3 E 1 This page is left blank intentionally...
Page 104: ......
Page 204: ...Contents Documents R S FSP 1164 4556 82 I 5 2 E 1 This page is left blank intentionally...
Page 208: ...Spare Parts R S FSP 1164 4556 82 5 4 E 2 This page is left blank intentionally...
Page 209: ...1164 4556 82 5 5 E 2 Spare Parts List Mechanical Drawings...
Page 222: ...List of R S FSP parts including spare parts R S FSP 1164 4556 82 5 18 E 2...
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ...11164 4556 82 5 19 E 2 Block Circuit Diagram...
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......