E10190-000816
FOS
Page 9
F)
Immediately after applying the adhesive, a Mylar or high temperature adhesive
tape (a "scotch" tape may also be used if curing temperature is low) may be
installed over the adhesive in the gauge sensitive region in order to keep the
gauge in good contact with the specimen. Do not apply excessive pressure over
the micro-capillary while applying the tape (this tape is removed after curing).
Follow the recommended cure schedule given by the adhesive manufacturer.
Note that the residual compression strain may increase with increasing curing
temperature. Post-curing the installation for two hours at least 15
C above the
maximum operating temperature with all tapes removed will provide essentially
creep-free performance.
Reminder: The key for successful strain gauge bounding is to have a good
contact between the gauge and the specimen to measure.
AE-10 epoxy
Side view
Side view
Risk of air bubbles
DO
DON'T
Mylar or Scotch
adhesive tape