4.
Review the calibration concentration description
Item
Steps
Display
1
Turn on the power, the screen shows
LOGO
2
Curve menu appears after 2 seconds
1 : H A C H < <
2 : L o v i b o n d
3
Press
button and
button
to select
the User curve
(
User-1 ~ User-6
),
”<<”
symbol represents the position of the
option, Press
button to enter the
Curve range menu
5 : U s e r - 2
6 : U s e r - 3 < <
4
Press
button to select 150
、
1500 or
15000.
“
<
”symbol represents the
position of the option, Press
button to
enter the Work menu
U s e r - 3 - 1 5 0 0
1 5 0 1 5 0 0 < 1 5 0 0 0
5
Select “
Info”
option, Press
button to
enter the User calibration data
U s e r - 3 - 1 5 0 0
S t a r t I n f o < C A L
6
Calibration result:
OK: Calibration parameters are correct.
NG: Calibration parameters are in error,
please go back to CAL and check again
U s e r - 3 - 1 5 0 0
C A L p o i n t = 2 O K <