Error messages
89
E
n
g
lis
h
Solution
Turn the monitor off and remove the test strip. Repeat the measurement
using a new test strip and blood taken from a new puncture site at the tip
of another finger. Do not touch or remove the test strip when a test is in
progress.
Note on Error 7 “Measurement”
: The CoaguChek XS PT test strip may
be used for patients under a combination therapy of oral anticoagulants
plus heparin injections. For maximum heparin concentrations which do
not interfere with the test, please refer to the package insert. Under no cir-
cumstances, however, should heparinized capillary tubes be used for
sample application. If capillaries are used, please use only the dedicated
CoaguChek capillary tubes. Be sure to apply the blood drop to the test
strip within 15 seconds of lancing the fingertip.
This error message (7) can occur in rare cases, i.e. in patients
with long coagulation times (> 8 INR, < 5% Quick). If this error
message appears again when you repeat the test, your results
must be checked using another method. Please contact the
doctor treating you immediately.
ccxs_user_EN.book Seite 89 Montag, 11. Juli 2005 7:11 19
Summary of Contents for CoaguChek XS System
Page 1: ...CoaguChek XS User s Manual Manuale per l uso ...
Page 4: ...The CoaguChek XS System 4 ...
Page 8: ...8 ...
Page 16: ...The CoaguChek XS monitor 16 The CoaguChek XS monitor G D A B E C F I H H ...
Page 42: ...Monitor setup 42 ...
Page 105: ...CoaguChek XS Manuale per l uso ...
Page 108: ...Il sistema CoaguChek XS 4 ...
Page 112: ...8 ...
Page 120: ...Lo strumento CoaguChek XS 16 Lo strumento CoaguChek XS G D A B E C F I H H ...
Page 146: ...Configurazione dello strumento 42 ...