![Roche CoaguChek XS System User Manual Download Page 70](http://html.mh-extra.com/html/roche/coaguchek-xs-system/coaguchek-xs-system_user-manual_1472287070.webp)
Cleaning
70
Cleaning the test strip guide
Check the test strip guide regularly for signs of soiling. If the test strip
guide has become soiled with blood or any other material, you must clean
this area.
1
Remove the measurement
chamber cover by pressing
upwards from the front (e.g.
using your thumb-nail).
You
can rinse the detached meas-
urement chamber cover (sep-
arately from the monitor) with
water.
2
Clean the easily accessible
white areas with a lint-free
swab or a moistened cotton
bud.
Make sure that
no
liquid
enters the monitor. Do not
insert any objects in the test
strip guide. Doing so may
damage the electrical con-
tacts behind the test strip
guide.
ccxs_user_EN.book Seite 70 Montag, 11. Juli 2005 7:11 19