13 ISE operation
Cedex Bio System
Replacing electrodes
Replacing electrodes
There are up to three ion-selective electrodes and one ISE Reference Electrode on the
ISE unit. The replacement procedure is the same for all of them.
As part of replacing electrodes, the
Electrode Service
maintenance action and a main
calibration need to be performed.
Replacing an electrode takes about 15 minutes.
Tools and materials required
o
ISE Calibrator indirect/urine (on board)
o
ISE Reference Solution (on board)
o
ISE Deproteinizer
o
ISE Etcher
o
Activator
o
ISE Solution 1
o
ISE Solution 2
o
ISE Sodium Electrode (if you intend to replace this type of electrode)
o
ISE Potassium Electrode (if you intend to replace this type of electrode)
o
ISE Reference Electrode (if you intend to replace this type of electrode)
o
ISE Dummy Electrode (if you intend to replace this type of electrode or if you no
longer intend to use one of the ion-selective electrodes)
Roche recommend replacing electrodes when their expiration date is reached.
For information on the stability of the electrodes, see their package inserts.
Make sure that you have read and understood section
The following warning messages in particular are relevant:
o
Injury through reagents and other working solutions
o
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...