prietest TOUCH
USER MANUAL
Version No.: 2.622A
Effective Date: July, 2010
Page 36 of 52
9.5. DIFFERENTIAL
Differential of Sample and Sample Blank is taken. Concentration is calculated either from the factor fed by
using standard.
PROGRAMMING / ADDING a NEW TEST (Refer chapter 8)
DIFFERENTIAL
Mode screen before programming would look like the above screen
Refer chapter11 for entering the test parameters .On completion, touch “SAVE” to save the programmed test.
Recalling /Running (Operating) a pre-programmed test. (Refer chapter 8)
For example:
.
Name
: Test Name,
Mod:
DIFF
36.45
Pri.
: 340 ,
Sec. :
0
ESC
Temp
: 37C ,
KF
: 1.000
EDIT
Vol:
500ul
,
Unit:
mg/dl
RUN
Lag :
5
, Read:
10
QCR
Blank:
Y,
QC
: N,
Normal:
Y
STD:
N,
Factor:
1746
Limit Set:
To change values press EDIT
Name
: ,
Mod:
DIFF
36.42
Pri. : 340
,
Sec. :
0
ESC
Temp
: 37C ,
KF
: 1.000
Vol:
300ul
,
Unit:
No –Unt
ADD
Lag :
0
, Read:
0
Blank:
N,
QC
: N,
Normal:
N
SAVE
STD:
N,
Factor:
0.000
Limit Set:
DIFFERENTIAL
Mode screen (After recalling a saved test.)
To change values, touch”
EDIT
”
To run the programmed test, touch “
RUN
”
To print, touch
“PRINT”
(Printing is optional)
To escape, touch “
ESC
”
Test Name:
ESC
PID
RUN
Aspirate Reference
Press ASP Switch to sip
At this stage,
the analyser will wait till the temperature of
cuvette reaches the set temperature. The instrument prompts
for REFERENCE, and prompts the user to place the reference
solution (distilled water) below the tube and to press the
aspiration switch. The instrument aspirates the solution and
automatically adjusts the reference.
(Note :To abort temperature setting, touch “
ESC
” once)
(In cuvette mode when the instrument prompts reference “close
the lid and Press RUN” before inserting sample cuvette)
Test Name:
ESC
PID
RUN
Wash
RunQC
Read Sample Blank
The user has to place the “sample blank” solution below the
tube and press the aspiration switch. The instrument aspirates
the sample, displays the “sample blank” absorbance.
(Note :At this stage the operator can enter the Patient
Identification. To enter Patient ID, touch”
PID
”. Entering PID is
optional. )
Test Name:
ESC
PID
RUN
Wash
RunQC
Aspirate Sample.
Press ASP Switch to sip
The user has to place the “sample” solution below the
tube and press the aspiration switch. The instrument
aspirates the sample, displays the “sample
absorbance” and computes the result.