prietest TOUCH
USER MANUAL
Version No.: 2.622A
Effective Date: July, 2010
Page 25 of 52
To change the mode to Flowcell, touch on “
cuvette
”
and then touch “
ESC
” to proceed
In the above screen, the analyser is in Flowcell Mode.
To change the mode to cuvette, touch on “
Flowcell
”
and then touch “
ESC
” to proceed
8.3.3 WASH F.C
Wash Flow cell is used to wash the flow cell .Selecting “
Wash F.C
” will allow the aspiration of water or cleaning solution
in the Flowcell.
8.3.4 Mode:
To Change Mode: Flow cell or Cuvette
It is possible with priete st TOUCH
to use either the Flowcell or cuvettes to execute readings. The
analyser is factory fitted with a Flowcell in the optical block. To operate the analyser with cuvettes
select “UTILITIES” in Main Menu and change the mode accordingly.
Flowcell Mode Cuvette Mode
.
In Flowcell mode, automatic pump calibration is implemented. So while doing the pump calibration, only
enter the exact measured residual volume out of 1000 µl. Don’t enter any less or higher value than the
residual volume. Refer to 8.3.5 for Pump Calibration.
36.40
ESC
MAIN MENU
1. Test Operations
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Communication
UTILITIES Ver (Version No) 36.68
ESC
1. Date.
2. Password
3. Wash F.C. 4. Flowcell
5. Pump Cal 6. Prog.Fil.
7. Printer On/Off : ON
8. Enter Clinic name
9. Lamp :ON
10. Communication: Serial
Enter (0-20) secs
in the Numeric screen
Touch “
ENTR
” ,upon completion.
Touch “
CLR
” ,to clear a wrong entry.
UTILITIES Ver (Version No) 36.64
ESC
1. Date.
2. Password
3. Wash F.C. 4. Flowcell
5. Pump Cal 6. Prog.Fil.
Numeric Value :
UUU
1 2 3 4 5 6 ENTR
7 8 9 0 . CLR
36.40
ESC
MAIN MENU
1. Test Operations
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Communication
UTILITIES Ver ( Version No) 36.64
ESC
1. Date.
2. Password
3. Wash F.C. 4. Flowcell
5. Pump Cal 6. Prog.Fil.
7. Printer On/Off : ON
8. Enter Clinic name
9. Lamp :ON
10. Communication: Serial
UTILITIES Ver ( Version No) 36.64
ESC
1. Date.
2. Password
3. Wash F.C. 4. Cuvette
5. Pump Cal 6. Prog.Fil.
7. Printer On/Off : ON
8. Enter Clinic name
9. Lamp :ON