8.3.5 Pump Cal:
The option of “
Pump Cal
” allows the operator to adjust the volume being aspirated through the
Flowcell.
The adjustment of the peristaltic pump has two functions:
To adjust the volume of liquid aspirated into the flow cell. The volume of liquid must be enough to
rinse Flowcell sufficiently, but not more than the reaction mixture;
To adjust the volume aspirated during a wash cycle
If the aspiration volume is too high, air will be sucked into the flow cell. If aspiration volume is too
low, not enough liquid will flush the Flowcell. Whenever this occurs it is necessary to adjust the
aspiration volume.
Step1: Go to utilities Step2: Select “Pump cal”
Step4: Enter residual volume.
Step 3: Take 1000 µl of de-ionized water in a
test tube and press aspiration switch.
36.40
ESC
MAIN MENU
1. Test Operations
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Communication
UTILITIES Ver (Version No) 36.46
ESC
1. Date.
2. Password
3. Wash F.C. 4. Flowcell
5. Pump Cal 6. Prog.Fil.
7. Printer On/Off : ON
8. Enter Clinic name
9. Lamp :ON
10. Communication: Serial
Enter residual volume in the Numeric screen
Touch “
ENTR
” ,upon completion.
Touch “
CLR
” ,to clear a wrong entry.
For example
: After pressing the ASP Switch, If
the volume remaining in the test tube is 250 µl
enter 250 in the above numeric screen and
touch
ENTR
.
The instrument is provided with a probe to
aspirate the de-ionized water /sample into the
flow cell. The push button (peristaltic
Switch/ASP switch) has to be pressed to
activate the peristaltic pump.
Aspirate 1000ul
Press ASP switch to sip
37.15
Enter Residual volume
Numeric Value: 250
1 2 3 4 5 6 ENTR
7 8 9 0 . CLR
Residual volume: 250
Calibration successful.
Same message was displayed on printer also.