4.2 Replacing rotating gobos
DANGER!
Install the gobos with the device switched off only.
Unplug from mains before!
1. Disconnect the fixture from mains and allow it to cool.
2. Remove the plastic cover of the fixture by loosening the 4 quarter-turn fasteners on the cover.
3. Gently pull up the gobo holder (3) from the rotation gobo wheel
.
4. Remove the spring lock (1) with an appropriate tool (e.g. small-bladed screwdriver) and remove it.
Do not touch the surface of the pattern of the glass gobo with bare fingers.
5. Remove the original gobo (2) and insert the new one (glazy side towards the light source).
The Robe gobo has a small position point (8) at its edge which has to aimed at the position point (4) on
the gobo holder (4). Insert the spring lock to secure correct gobo position in the gobo holder.
6. Insert
he gobo holder back under the distance slots (6, 7) into rotating gobo wheel in this way, that its
position point (4) has to aimed at a small toothlike projection (5) on the edge of the rotating gobo wheel.
Important!
When inserting the gobo holder back to the rotating gobo wheel, one of the adjacent gobo holder
has to be oriented according to the same rule, it means that its position point (4) has to aimed at
the toothlike
projection (5) on the edge of the rotating gobo wheel. You have keep both marks (4) and (5) side by side when
rotating the gobo wheel to the position allows inserting the gobo holder back.
5. Replace the plastic cover before applying power.
6. Use the menu Service to fine adjust replaced gobo (Service -> Calibration -> Calibrate effects -> R. Gobo
Index 1 ...R. Gobo Index 6).
Note. The magnet (9) of the gobo holder has the same function as the position point (4) on the rest of gobo
holders.