3 Recall to output
In any state, just press the buttons M1 M5 and then the according memories can
be recalled.
4 The Operation of Overcurrent Protection
Press and hold the button “OCP” for 3 seconds to enter the mode of OCP setting,
when the indicator “OCP SET” lights on. And the current values on both CH1 and
CH2 display the OCP values accordingly. By adjusting the current knob, the OCP
value can be changed. Press and hold the button “OCP” again for 3 seconds to
exit. Furthermore, press the button “OCP” to switch on the Overcurrent Protection
(OCP) mode and the indicator “OCP” is turned on; press the button “ OCP” again
to shut down the OCP mode and the indicator “OCP”will be turned off. When the
OCP mode is on, if the current value on the load or the setting current is more
than that in the OCP SET, the output will be cut off.
5 The Overvoltage Protection (OVP) Setting and Switching on
Press and hold the button “OVP”for 3 seconds to enter the mode of OVP setting,
when the indicator “OVP SET” lights on. And the voltage values on both CH1 and
CH2 display the OVP values accordingly. By adjusting the voltage knob, the OVP
value can be changed. Press and hold the button “OVP” again for 3 seconds to
exit. Furthermore, press the button “OVP” to switch on the Overvoltage Protec-
tion (OVP) mode and the indicator “OVP” is turned on; press the button “OVP”
again to shut down the OVP mode and the indicator “OVP” will be turned off.
When the OVP mode is on, if the voltage value on the load or the setting voltage is
more than that in the OVP SET, the output will be cut off.