3 Types of totalizers
Manual TRZ 03· EN08 2018 September, 10th
34
Notice
This design provides no options for connecting additional mechanical
equipment.
The totalizer has the following features:
•
LF pulse transmitter
- Standard:
reed-contact
- Alternatively:
inductive pulse transmitter
- Option:
up to two additional inductive LF pulse transmitters
•
HF pulse transmitters are optionally available,
pulse frequency approx. 100 Hz at Q
max
•
IP 65 Protection class
•
Universal readable
•
Totalizer unit and HF 1 pulse transmitter are easily replaceable on site.
Notice
The electrical connector must be covered by a cap closure or a connecting
plug! Otherwise moisture may ingress into the meter head.
The ENCO-08 output connector is placed in the
meter head. It is 3-pole and has the assignment:
1
– Vo (+)
2
– --- (not used)
3
– Vo (-)
4
– PE (protective earth)