59
As a policy of continual improvement, RMF reserves the right to alter the specification without prior notice.
201.028 REV 1
Date of Issue: 18 June 2018
10-17
Test Reference
Array of 16 packed characters
18
Test Duration
Unsigned integer
19
Test Format
20
Test Mode
21
Command
Unsigned integer
22-23
Test Interval
S
Unsigned 32 bit integer
24-25
Date/Time
Date
Unsigned 32 bit integer
26
Alarm Mode
Unsigned integer
27
Reserved
28
Faults
29
Reserved
30
Status
Unsigned integer
31
Status Flags
32
LED Level
Unsigned integer
33
Temperature
°C x100
34
RH
% x100
35
Peak Pulse
Unsigned integer
36
Test Completion
Unsigned integer
37
Flow Indication
Unsigned integer
38-39
Reserved
40-55
Particle Counts
Array of 8 32 bit integers
56-63
Result Codes
Array of 8 signed integers
64-71
Contamination Limit Upper
Array of 8 signed integers
72-79
Contamination Limit Lower
Array of 8 signed integers
80
Limit Water Upper
% x100
signed integer
81
Limit Water Lower
% x100
signed integer
82
Limit Temperature Upper
°C x100
signed integer
83
Limit Temperature Lower
°C x100
signed integer
84-85
Log Interval
Seconds
Unsigned 32 bit integer
86-87
Last Download
Date
Unsigned 32 bit integer
88
Language
Unsigned integer
89-116
Reserved
117-118
Calibration Due
Date
Unsigned 32 bit integer
119-120
Calibration Last
Date
Unsigned 32 bit integer
121
Reserved
122
Calibration LED Level Last
123
Calibration LED Level Initial
124
Reserved
Table 11A Modbus Register Map
11.1.1.2 Representations
Modbus Registers
- All quantities are represented using Modbus registers. Modbus registers are 16 bit (0-65535
decimal or 0-0xFFFF in hexadecimal notation).
Unsigned Integers
- These are simply single Modbus registers. Each can take values from 0 to 65535.
Summary of Contents for CMS 2
Page 1: ...User Manual CMS2...