- 43 -
SD-1GH
WARNING
Always perform both a zero adjustment and a span adjustment in that order. If they are performed in
a wrong order or only one of them is performed, accurate gas detection cannot be ensured.
WARNING
Be sure the detector is in a fresh air environment before performing a zero adjustment. If the
detector is not in a fresh air environment, a cylinder of zero air must be used to perform the zero
adjustment.
7-3. Performing a Calibration
Perform a zero adjustment and span adjustment as part of a complete calibration.
Materials:
•
Zero adjustment gas
•
Span adjustment gas
•
0.5 LPM fixed flow regulator
•
Humidifier tube
Detection Gas
Full Scale
Calibration Gas/
Concentration
Calibration
Value
C2H5OH (Ethanol)
2000 ppm IPA: 1000 ppm
1300 ppm
5000 ppm IPA: 2000 ppm
2500 ppm
C3H3N (Acrylonitrile)
100 ppm
Isobutane: 50 ppm
70 ppm
C5H10O2 (Propyl acetate)
1000 ppm IPA: 500 ppm
500 ppm
C6H6 (Benzene)
100 ppm
Isobutane: 75 ppm
35 ppm
500 ppm
Isobutane: 1000 ppm 300 ppm
C6H14 (Hexane)
500 ppm
Hexane: 360 ppm
360 ppm
3000 ppm Hexane: 1800 ppm
1800 ppm
CH2Cl2 (Methylene chloride) 500 ppm
Isobutane: 500 ppm
180 ppm
CH4 (Methane)
2000 ppm Methane: 1000 ppm
1000 ppm
EO (Ethylene oxide)
1000 ppm Hydrogen: 500 ppm
500 ppm
MEK (Methyl ethyl ketone)
500 ppm
Hexane: 1100 ppm
200 ppm
R404a
2000 ppm Isobutane: 250 ppm
1000 ppm
* Not all listed cylinders have RKI part numbers (see pg. 57 for cylinder part
numbers). Contact RKI if you need a cylinder that does not have a part number.
<Zero Adjustment "2-1">
Detection Mode
While in Detection Mode,
press and hold the
MENU/ESC key for three
seconds.
User Mode
1-0. ROM/SUM displays.
Scroll to the 1-3 Mode
Switching item.