- 37 -
SD-1GH
WARNING
•
Before starting the alarm test (transmission test), inform related sections so that they can
prepare for false abnormalities (external output signals and alarm contact). After the test is
completed, press MENU/ESC key to return to Detection Mode. (If the detector remains in Alarm
Test Mode, it automatically returns to Detection Mode in ten hours.)
•
In Alarm Test Mode, the gas concentration equal to the reading is output depending on the
Maintenance Mode setting. Take actions, if necessary, so that the signal receiver side can
prepare for false alarms.
<Alarm Test Mode "2-0">
Enter Maintenance Mode as described on page 35.
2-0
Press SET key.
2-0.0 Gas Test
2-0.1 Alarm Test
2-0.2 Fault Test
2-0.3 LED Test
2-0.4 Memory Test
Gas Test => P38
Alarm Test =>P39
Fault Test => P40
LED Test => P40
Memory Test => P41
To 2-0.0