47
4
Remove the old sensor.
5
Install the new sensor.
Be sure the sensor pins are lined
up with the board’s sockets.
6
Put the sensor cover back
on.
CAUTION
•
When installing the sensor, be sure the pins are lined up with the sockets before pushing in the
sensor. Ignoring this may bend and damage the sensor pins.
•
After the sensor is installed, always perform a calibration (zero adjustment and span adjustment).
•
Do not disassemble the sensor because it contains electrolyte. If contact occurs, rinse the area
immediately with a large quantity of water.
•
The sensor must be stored under normal temperature/humidity in a clean place away from direct
sunlight.