04 Series for CO(-H2)
Appendix A: Maintenance Mode • 87
3. Press and release POWER MODE to perform a default operation. The instrument asks if
you are sure you want to perform a default operation.
4. Press and release POWER MODE to perform a default operation. The instrument beeps
twice and returns to the
M.DEF
item in Maintenance Mode.
5. See “Entering Measuring Mode (START)” on page 87 to enter Measuring Mode.
Entering Measuring Mode (START)
1. While in Maintenance Mode, press AIR to scroll to
START
.
2. Press and release POWER MODE. The instrument begins its warmup sequence.
SURE ?
YES .MODE
NO .AIR
START