04 Series for CO(-H2)
Chapter 5: User Mode and Calibration • 53
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
CAL.INT
item in
the
CAL SET
menu.
5. See “Exiting the CAL SET Menu” on page 53 to return to User Mode.
CAL.EXPD
This item defines what indication is given during start up when calibration is due and
CAL.RMDR
is set to
ON
(factory setting).
CONFIRM
(factory setting): The CO-04C gives an indication at start up if calibration is past
due. Press and release AIR to continue without calibrating or press and release POWER
MODE to enter User Mode and perform a calibration.
CANT.USE
: The CO-04C gives an indication at start up that calibration is past due. Press and
release POWER MODE to enter User Mode and perform a calibration. Pressing AIR has no
effect. A successful calibration must be performed in order to use the instrument.
NONE
: The CO-04C gives an indication at startup that calibration is past due. It is not
necessary to acknowledge the indication. If desired, press POWER MODE to enter User
Mode and perform a calibration. The warm-up sequence continues on its own.
1. After entering the
CAL SET
menu, press AIR to scroll to
CAL.EXPD
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
CAL.EXPD
item
in the
CAL SET
menu.
5. See “Exiting the CAL SET Menu” on page 53 to return to User Mode.
Exiting the CAL SET Menu
1. While in the
CAL SET
menu, press AIR to scroll to
ESCAPE
.
2. Press and release POWER MODE to return to the
CAL SET
item in User Mode.
3. See “Entering Measuring Mode (START)” on page 64 to enter Measuring Mode.
Setting Bump Test Parameters (BUMP.SET)
The
BUMP.SET
menu has 5 items:
SETTING
,
BP.RMDR
,
BP.INT
,
BP.EXPD
, and
ESCAPE
.
1. While in User Mode, press AIR to scroll to
BUMP.SET
.
CAL .EXPD
BUMP .SET