04 Series for CO(-H2)
Chapter 5: User Mode and Calibration • 51
4. Press and release POWER MODE. The Auto Cal screen for the gas assigned to Cylinder
A displays.
5. Use AIR to scroll to
CYL SEL
.
6. Press and release POWER MODE. CO displays.
7. Use AIR to scroll to the gas whose cylinder assignment you want to change.
8. Press and release POWER MODE. The current setting flashes. In the example below, CO
is selected.
9. Use AIR to change the cylinder assignment. The choices are A, B, C, D, and E.
10. Press and release POWER MODE to save the change. The cylinder assignment stops
flashing and the unit returns to the gas selection screen.
11. Repeat steps Step 7 - Step 10 for the other channel, if necessary.
12. Use AIR to scroll to
ESCAPE
.
13. Press and release POWER MODE. The instrument returns to the
CYL SEL
menu item in
the Auto Cal Menu.
14. Press AIR to scroll to
ESCAPE
.
15. Press and release POWER MODE. The instrument returns to the
A-CAL
menu item in the
GAS CAL
menu.
16. See “Exiting the GAS CAL Menu” on page 52 to return to User Mode.
CO
50
A-CAL
ppm
CO
50
CYL A
ppm
CYL SEL
CO
CYL SEL
ppm
AIR
AIR
AIR
CO
CYL SEL
ppm
H2
CYL SEL
ppm
ESCAPE
CO
A
CYL SEL
ppm