VOC Pro Transmitter • 63
9. Turn the regulator knob counterclockwise to open the regulator.
10. Ensure that the gas is flowing and press the MENU button to begin calibrating the sensor,
the unit will automatically begin the calibration process, the amount of time on the timer
will vary based on the gas type. During calibration, the display will show a countdown of
the time remaining until the process is complete.
NOTE:
Once the calibration countdown has started, the process cannot be stopped without
disconnecting the power from the unit.
11. When calibration is complete, remove the calibration cup from the sensor housing and
reinstall the rain guard.
NOTE:
If the sensor responds extremely slow, or does not respond to the applied gas, it may
indicate a failed sensor element. The sensor element will need to be replaced before
completing the zero and calibration process.
12. Turn the regulator knob clockwise to close the regulator.
13. Press the MENU button until the normal operating screen appears.