26 • VOC Pro Transmitter
Powering On
Powering on the device begins the operation of the unit, automatically initiating the system
start-up cycle and 1-minute warmup period. The VOC Pro will be in normal operating mode
at the completion of the system start-up.
1. Press the ADD button once to turn the unit on.
Manual Reset for Activated Latching Alarms
Relay alarms set to latching will not deactivate until the alarms are manually reset at the
device. This includes Alarm 1 and Alarm 2 indicator LEDs and optional wired relays. When
latching alarms have been activated, refer to the following instructions for how to manually
deactivate the alarms on your device:
1. Verify that the gas level reading is below the alarm level setting.
2. Press the MENU button to deactivate latching alarm(s).
NOTE:
The gas level reading MUST be below the alarm level setting before the alarm can be
deactivated. Press the MENU button ONLY once to deactivate the latching alarm(s).
Pressing the MENU button more than once will activate and open the Operation
Settings menu. The alarms will NOT activate, even in the presence of gas, until you
have exited the menu mode for approximately 1 minute.
Faults
In the event of a device failure, the unit will alternate between the normal operating screen and
a fault screen on the display, in 5 second intervals, until the fault has been cleared, or is
corrected. The fault code, located in the bottom-left corner of the display, appears on both
screens. The unit continuously registers that the system is in fault. When the fault is corrected,
the unit will return to normal operating mode.
For a list of the fault codes and warning symbols of the VOC Pro, and their associated
meaning, see page 44.