4. Locate the fuse that needs to be replaced.
5. Pull the old fuse straight out.
6. Align the pins on the replacement fuse with the sockets in the fuse holder.
7. Push the new fuse into the fuse holder.
8. Place the internal system back into the enclosure, matching each mounting post to its
corresponding eyelet anchored within the base of the enclosure.
9. Using the thumbscrews, gently push to seat the internal system into the mounting posts.
NOTE:
The thumbscrews on the VOC Pro function
ONLY
as thumb-holds for ease in
removal of the internal system from the base of the enclosure. Do
NOT
attempt to
loosen or tighten the thumbscrews when opening or closing the enclosure.
10. Verify that the sealing ring, seated at the threaded opening of the device enclosure, is
correctly in place.
11. Secure the enclosure lid back onto the enclosure.
WARNING: When securing the lid onto the device, tighten the enclosure lid by hand
ONLY. Overtightening of the lid by use of hand-tools could result in damage to
the O-ring, potentially compromising the moisture seal, resulting in an unsafe
environment.