6 • 35-3010A-11 Sample-Draw Detector
Description
This section describes the components of the
35-3010A-11 sample draw detector
. The sample
draw detector consists of the housing, flow system, and detection system.
Figure 1: Sample Draw Detector Component Location
Housing
The sample draw detector’s fiberglass housing is weather- and corrosion-resistant. It is
suitable for installation where general purpose equipment is in use. The housing door is
hinged on the left side and is secured by two latches on the right side.
Four mounting feet are attached to the back of the housing (one at each corner). Use the
mounting feet to install the housing to a vertical surface. Use the two conduit hubs on the
bottom of the housing to make wiring connections.
An aluminum subpanel is mounted to the interior of the housing. The sample draw
detector’s internal components are mounted to the subpanel.
Amp 2
Amp 1
Combustible G as
and H2S Flow B lock
IR CO2
Flow B lock
IR CO2
Sensor
Preamp
Circuit Board
IR CO2 Transmitter
Main Circuit B oard
Pump
Hydr ophobic Filter
(A croPak)
Ex haust Fitting For 1/4" O D Rigid Tubing
Charcoal Filter
Combustible G as S ensor
Dummy P lug for O xygen S ensor
Dummy P lug for CO Sensor
H2S S ensor
3/4 C onduit H ub, 2X
Inlet Fitting For 1/4" OD R igid Tubing
Reset Switch