35-3010A-11 Sample-Draw Detector • 25
3.
Connect the sample tubing from the regulator to the inlet line at or near the INLET
fitting.
4.
Allow the gas to flow for 1 minute.
5.
Verify a voltmeter reading of 100 mV ± 2 mV at the AMP 2 test points for the H
2
S
channel as described in the Preparing for Calibration section above.
6.
If necessary, use a small flat-blade screwdriver to adjust the zero pot for the H
2
S
channel until the voltmeter reading is 100 mV ± 2 mV.
7.
Verify a voltmeter reading of 100 mV ± 2 mV at the IR CO
2
test points as described in
the Preparing for Calibration section above.
8.
If necessary, use a small flat-blade screwdriver to adjust the zero pot for the CO
2
channel until the voltmeter reading is 100 mV ± 2 mV.
9.
Disconnect the sample tubing from the inlet line.
10. Unscrew the regulator from the zero air cylinder. Leave the sample tubing connected
to the regulator.
Setting the Response Reading
1.
Screw the regulator into the 4-gas calibration cylinder.
2.
Connect the sample tubing from the regulator to the inlet line at or near the sample
draw detector’s INLET fitting.
3.
Allow the calibration gas to flow for 1 minute.
4.
Check the mV output on the AMP 2 test points for H
2
S and verify that the reading
matches the response reading (±2 mV) you determined earlier.
5.
If necessary, use the AMP 2 span pot to adjust the reading to match the correct
response reading.
6.
Disconnect the sample tubing from the sample draw detector’s inlet line.
7.
Unscrew the regulator from the 4-gas calibration cylinder.
8.
Screw the regulator into the CO
2
calibration cylinder.
9.
Connect the sample tubing from the regulator to the inlet line at or near the sample
draw detector’s INLET fitting.
10. Allow the calibration gas to flow for 1 minute.
11. Check the mV output on the IR CO
2
test points and verify that the reading matches
the response reading (±2 mV) you determined earlier.
12. If necessary, use the span pot on the IR CO
2
transmitter to adjust the reading to match
the correct response reading.
13. Disconnect the sample tubing from the sample draw detector’s inlet line.
14. Unscrew the regulator from the CO
2
calibration cylinder.
NOTE:
For convenience, leave the regulator connected to the sample tubing.
Returning to Normal Operation
1.
Remove the voltmeter leads from the test points.
2.
Reconnect the incoming sample line.