background image

 

12 • 35-3001A-09 Oxygen/Hydrogen Sulfide Sample-Draw Detector

9. If shielded cable is used, connect the cable’s drain wire to an available chassis (earth) ground at 

the controller. RKI controllers typically have a ground stud that can be used to ground the 
cable’s drain wire.

Start Up

This section describes procedures to start up the sample-draw detector and place the sample-draw 
detector into normal operation.

Introducing Incoming Power

1. Complete the installation procedures described earlier in this manual.

2. Verify that the wiring is correct and secure. Refer to the controller operator’s manual for 

connections at the controller.

3. Turn on or plug in the power to the controller, then turn on the controller.
4. Verify that the sample-draw detector’s Pilot LED is on.
5. Verify that the controller is on and operating properly. Refer to the controller operator’s 

manual.

6. Verify that the flowmeter indicates a flow rate of approximately 1.0 SCFH. If necessary, use 

the sensor flow control valve to adjust the flow rate. Turn the valve’s knob clockwise to 
increase the flow and counterclockwise to decrease the flow.

NOTE:

The following step tests for leaks in the sample line. This test will cause a low flow 
condition at the sample-draw detector and a fail condition at the controller. Be sure to put 
the controller into its calibration program or disable external alarms before performing 
this test.

7. Verify that the incoming sample line is not leaking. To test the sample line, press and hold the 

reset switch and plug the open end of the sample line with your thumb. If the flowmeter ball 
drops to the bottom of the flowmeter, the incoming sample line is not leaking.

8. Remove your thumb from the sample line, release the reset switch, and verify the flowmeter 

returns to a normal flow rate.

9. Enable alarms or place the controller in normal operation.

CAUTION:

Allow the sample-draw detector to warm up for 5 minutes before you continue with 
the next section, “Setting the Zero Reading.”

Setting the Zero Reading

CAUTION:

If you suspect the presence of hydrogen sulfide or an abnormal oxygen concentration 
in the monitoring environment, use the calibration kit and the zero air calibration 
cylinder to introduce “fresh air” to the sensor and verify an accurate zero setting. 
See “Calibration, Oxygen Sensor” on page 21 an
d “Calibration, H

2

S Sensor” on 

page 23 for instructions on using a zero air calibration cylinder for setting the zero 
reading.

1. Verify that the sample-draw detector is sampling a fresh air environment (environment known 

to be free of toxic gas and of normal oxygen concentration, 20.9%).

2. Open the housing door.

Summary of Contents for 35-3001A-09

Page 1: ...www rkiinstruments com 35 3001A 09 Oxygen Hydrogen Sulfide Sample Draw Detector Operator s Manual Part Number 71 0311RK Revision A Released 8 8 17...

Page 2: ...eath Periodic calibration and maintenance of the detector is essential for proper operation and correct readings Please calibrate and maintain this detector regularly Frequency of calibration depends...

Page 3: ...l or replacement costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESE...

Page 4: ...ing the Zero Reading 12 Maintenance 13 Preventive Maintenance 13 Troubleshooting 15 Replacing Components of the Sample Draw Detector 16 Adjusting the Low Flow Setting 19 Removing the Particle Filter s...

Page 5: ...he risk of personal injury Be sure to maintain and periodically calibrate the 35 3001A 09 as described in this manual Table 1 Specifications Target Gases and Detection Range Oxygen 0 25 volume Hydroge...

Page 6: ...H2S Amplifier H2S Sensor inside block Pump Pump Connector Oxygen Sensor Terminal Strip Mounting Foot 4X Oxygen Amplifier S OXY T O XIC SIG PWR B G H2S Flow Block Oxygen Sensor S AMP AMP 2 S OXY R Fai...

Page 7: ...but it is not factory installed If the particle filter is installed directly to the inlet fitting the tubing stub must be used If the particle filter is installed somewhere else like at the end of th...

Page 8: ...e Installation section on page 8 for all wiring procedures related to the sample draw detector Pump Connector The pump connector is the two point connector below the oxygen terminal strip in the upper...

Page 9: ...The pump reset switch is located to the left of the status LEDs When a low flow condition occurs the pump will be shut off To reset the low flow condition and start the pump again press and hold the p...

Page 10: ...clockwise to decrease the fresh air output CAUTION The amplifier includes an additional pot It is factory set Do not adjust it Controller and Detector Terminal Strips The controller terminal strip and...

Page 11: ...adjustment screw and adjust the amplifier s zero output during the start up and calibration procedures Turn the adjustment screw clockwise to increase the zero output and counterclockwise to decrease...

Page 12: ...ollowing when you select the mounting site Is there enough room to open the housing door and make wiring and tubing connections at the bottom of the housing Make sure there is sufficient room to perfo...

Page 13: ...fitting components 4 Connect a length of sample tubing to the other side of the particle filter and route it to the sampling area 1 4 O D rigid polypropylene Teflon or flexible polyurethane tubing ma...

Page 14: ...priate insert to seal the connection between the tubing and the inlet fitting 3 Hand tighten the nut on the exhaust fitting so the ferrules clamp on the tubing If the tube nut is removed see Figure 7...

Page 15: ...s manual and the controller detector head specification sheet for the 35 3001A 09 for detector terminal strip connections specific to the controller Figure 6 Wiring the Sample Draw Detector to a Cont...

Page 16: ...ondition at the sample draw detector and a fail condition at the controller Be sure to put the controller into its calibration program or disable external alarms before performing this test 7 Verify t...

Page 17: ...ace components of the sample draw detector adjust the low flow setting and remove the particle filter s tubing stub Preventive Maintenance This section describes a preventive maintenance schedule to e...

Page 18: ...ulator into the hydrogen sulfide calibration cylinder 2 Connect the calibration tubing from the demand flow regulator to the inlet fitting Gas will begin to flow 3 After approximately one minute verif...

Page 19: ...the correct flow rate with the sensor flow control valve 2 If you cannot set the correct flow rate check the sample lines for obstructions or kinks 3 Verify that the sensor wiring is correct and secur...

Page 20: ...controller 3 Open the housing door of the sample draw detector 4 Loosen the two screws on the oxygen terminal strip and remove the lugs and wires from the terminal strip 5 Unscrew and remove the two s...

Page 21: ...power to the controller 3 Open the housing door of the sample draw detector 4 Remove the detector and controller plug in terminal strips by grasping the sides of the strips with your fingers Let the t...

Page 22: ...sensor off of the flow block 6 Disconnect the sensor from the H2S amplifier 7 Plug the H2S sensor into the new H2S amplifier 8 Place the sensor and amplifier in the H2S sensor cavity 9 Secure the H2S...

Page 23: ...stall the tubing routed to the sampling area 4 If the particle filter is installed at the end of the inlet line 1 Remove the particle filter from the tubing stub 2 Install the new particle filter onto...

Page 24: ...mes factory installed in the particle filter It is used for connecting the particle filter to the inlet fitting If you have installed the particle filter and no longer want it installed you will need...

Page 25: ...an every 3 months may be necessary Calibration Oxygen Sensor This section describes how to calibrate the oxygen sensor in the sample draw detector It includes procedures to prepare for calibration set...

Page 26: ...H2S in nitrogen to zero the oxygen sensor 2 Connect the sample tubing from the regulator to the inlet line 3 Allow the sample draw detector to draw the calibrating sample for 1 minute 4 Verify a volt...

Page 27: ...ple Output mV calibrating sample fullscale X 400 100 For example with a test sample of 25 ppm and a fullscale setting of 100 ppm the correct output is 200 mV 200 mV 25 100 X 400 100 Setting the Zero R...

Page 28: ...reading 6 Disconnect the sample tubing from the inlet line 7 Unscrew the regulator from the hydrogen sulfide calibration cylinder Returning to Normal Operation 1 Wait approximately one minute to allo...

Page 29: ...Particle filter 57 0140RK H2S amplifier 57 1064RK 11 Oxygen amplifier 71 0311RK Operator s Manual 35 3001A 09 Sample Draw Detector 81 0076RK 01 Zero air calibration cylinder 34 liter steel 81 0076RK 0...

Reviews: