IM100CZ04-E4

 

7

2. 

Set the value obtained in 1 above in the Gain setting section of 
the instrument *. 

 

Example:  When the e

0

 

 1.500 SUS630 diaphragm material’s

 

pressure sensor is used at 230 

C. 

 

e

= [1 + 0.13 

 10

-3

 

 (230

150)] 

 1.500 

e

= 1.516 

Set 1.516 to the gain setting portion of the instrument. 

 

For PCT-300, use the Gain Setter. 
For PG500, refer to Gain setting in the Setup setting mode in the PG500 
Operation Manual (IMR02F02-E

).

 

For HA430/HA930, refer to the Gain setting in Function block F21 in the 
Engineering mode in the HA430/HA930 Quick Operation Manual 
(IMR01N16-E

).

 

 
 

6.2 Correcting Indication Error due to the
 

Connection Cable Length 

RKC Instrument’s resin pressure sensor is calibrated for the 
standard cable length (5 m). Therefore, if the total length of the 
pressure sensor connection cable connected to the intrinsically safe 
circuit and non-intrinsically safe circuit side is other than 5 m, the 
resin pressure sensor indication value can be corrected using the 
instrument’s gain setting. 

 

 

Correction procedure 

1. 

The rated output after the correction is calculated from the 
following correction equation and correction factor. 

e

1

 

= e

0

’/[1 + K 

 (L 

 5)] 

e

0

’: Rated output of the resin pressure sensor

 

e

1

’: Rated output after the correction 

K (Correction factor): 
 1.96 

 10

-4

/m (Standard specification type) 

 1.40 

 10

-4

/m (Explosionproof specification type) 

L:  Cable total length (m) 
 

[However, when using RKC Instrument’s standard cable.] 

 

2. 

Set the value obtained in 1 above in the Gain setting section of 
the instrument *. 

 

Example: 
 

Cable total length (L) 

 10 m 

 

Rated output of the resin pressure sensor (e

0

’) 

 1.500 

 

e

1

 

= 1.500/[1 + 1.96 

 10

-4

 

 (10

5)] 

e

1

’ = 1.499 

Set 1.499 to the gain setting portion of the instrument. 

 

For PCT-300, use the Gain Setter. 
For PG500, refer to Gain setting in the Setup setting mode in the PG500 
Operation Manual (IMR02F02-E

). 

For HA430/HA930, refer to the Gain setting in Function block F21 in the 
Engineering mode in the HA430/HA930 Quick Operation Manual 
(IMR01N16-E

).

 

 

 

 

6.3 Correcting Indication Error when the
 

Cable of Another Company is Used 

The nominal cross-sectional area of our cable conductor is 0.5 mm

2

If using a cable from another company, the resin pressure sensor’s 
indication value can be corrected by correcting the gain setting of 
the instrument. 

 

 

Correction procedure 

1. 

The rated output after the correction is calculated from the 
following correction equation and correction factor. 

e

1

’ = e

0

’/[1 + 0.5/S 

 K 

 (L 

 5)] 

e

0

’: Rated output of the resin pressure sensor

 

e

1

’: Rated output after the correction 

S:  Conductor nominal cross-sectional area (mm

2

K (Correction factor): 
 1.96 

 10

-4

/m (Standard specification type) 

 1.40 

 10

-4

/m (Explosionproof specification type) 

L:  Cable total length (m) 

 

2. 

Set the value obtained in 1 above in the Gain setting section of 
the instrument *. 

 

Example: 
Cable total length (L) 

 10 m 

Conductor nominal cross-sectional area (S) 

 0.75 mm

2

 

Rated output of the resin pressure sensor (e

0

’) 

 1.500 mV/V 

 

e

1

’ = 1.500/[1 + 0.5/0.75 

 1.96 

 10

-4

 

 (10

5)] 

e

1

’ = 1.499 

Set 1.499 to the gain setting portion of the instrument. 

 

For PCT-300, use the Gain Setter. 
For PG500, refer to Gain setting in the Setup setting mode in the PG500 
Operation Manual (IMR02F02-E

).

 

For HA430/HA930, refer to the Gain setting in Function block F21 in the 
Engineering mode in the HA430/HA930 Quick Operation Manual 
(IMR01N16-E

).

 

 
 

When anti-explosion specifications are required, use a 
cable with an allowable capacitance between the cable 
wires of 0.1

F or less and an allowable inductance of    

0.6 mH or less. 

 

The allowable inductance could be exceeded if the cable is 
wound, so do not wind the cable when using it. 

 

 

6.4  Correcting Indication Error due to the
 Safety 

Barrier 

A pressure indication error caused by the dispersion of our RZB-001 
internal resistance value is within about 1 % of span. However, when 
this error needs to be lessened further, make the correction, if 
necessary. For this correction, the barrier correction factor B is used. 
No correction is required when the barrier correction factor B is 
“1.000.” 

 

 

The barrier correction factor B is described on the 
nameplate of the RZB-001. 

 

 

Correction procedure 

1. 

The rated output after the correction is calculated from the 
following correction equation and correction factor. 

e

1

’’

 

= B 

 e

0

’’ 

e

0

’’:  Rated output of the resin pressure sensor

 

e

1

’’:  Rated output after the correction 

B:  Barrier correction factor 

 

2. 

Set the value obtained in 1 above in the Gain setting section of 
the instrument *. 

 

Example: 
Barrier correction factor (B) 

 1.001 

Rated output of the resin pressure sensor (e

0

’’) 

 1.500 mV/V 

Rated output after the correction

 

 

e

1

’’ 

 

e

1

’’

 

= 1.001 

 1.500 

e

1

’’ = 1.502 

Set 1.502 to the gain setting portion of the instrument. 

 

For PCT-300, use the Gain Setter. 
For PG500, refer to Gain setting in the Setup setting mode in the PG500 
Operation Manual (IMR02F02-E

).

 

For HA430/HA930, refer to the Gain setting in Function block F21 in the 
Engineering mode in the HA430/HA930 Quick Operation Manual 
(IMR01N16-E

).

 

 

 
 

 
 
 
 

Summary of Contents for CZ-100P

Page 1: ...wer is turned on again To prevent instrument damage as a result of failure protect the power line and the input output lines from high currents with a suitable overcurrent protection device with adequ...

Page 2: ...Diaphragm section material S SUS630 Standard H HASTELLOY C Optional 3 Diaphragm surface treatment N Standard K CERAMIC Kanigin plate Optional 4 Intrinsically safe N Standard For non explosionproof spe...

Page 3: ...he slanting direction Fig 1 B to the horizontal direction Fig 1 C in this order In this case take measures of 1 and 2 if necessary by checking the sensor outer cylinder surface temperature To the rele...

Page 4: ...llow loose fitting Z 1024 or Z 1024A type 0 to 1 MPa Z 1024 or Z 1024A type 24 0 0 2 3 1 5 Pressure sensor installation 1 Make sure the mounting hole is correctly machined If installing the pressure s...

Page 5: ...maximum available current of 8 A Use the solderless terminal appropriate to the screw size Screw size M3 6 Recommended tightening torque 0 4 N m 4 kgf cm Specified solderless terminals With isolation...

Page 6: ...de of PCT 300 refer to drawing on page 6 to 2 Gain setting for CZ 100P LLA Z 1024A type Set the RATED OUTPUT Rated output mV V value specified on the label of the pressure sensor at Gain setting in th...

Page 7: ...wing correction equation and correction factor e1 e0 1 0 5 S K L 5 e0 Rated output of the resin pressure sensor e1 Rated output after the correction S Conductor nominal cross sectional area mm2 K Corr...

Page 8: ...on is fluctuated Value different from the sensor rated output set to the rated output setter for the PCT 300 The diaphragm deteriorated deformed or damaged The sensor exposed to hot or cold wind Some...

Page 9: ...aphragm deformed or damaged as the push rod pushes the diaphragm from the inside as a result of the deformation of the sensor flange by excessive tightening torque 9 EXPLANATION OF EACH TERMS Term Exp...

Page 10: ...0 5 As the input side of bridge resistance the 350 5 type is also available This type is interchangeable with the 374 10 type Temperature characteristics Maximum diaphragm temperature 400 C Maximum st...

Page 11: ...Linearity Within 0 01 of span Noise Within 0 1 p p of span 0 1 to 10 Hz Response 10 Hz 100 Hz transfer type Factory shipment 10 Hz Power supply 90 to 264 V AC Including power supply voltage variation...