Operator
’s Guide
Inspector 5000
™ Auto Optic
003-4500
RJS, Minneapolis, MN
27
3 MIL
0.003 inches
*5 MIL
6 MIL
0.005 inches
(for Europe and Japan)
0.006 inches
10 MIL 0.010 inches
20 MIL 0.020 inches
See Appendix I for an explanation of how to select
the correct aperture size
The next selection is Wavelength:
This selection allows you to select one of two light
wavelengths which are Red (660nm) or Infrared (925nm -
invisible). The default it RED and Infrared is used by very
few customers.
The next selection is Auto Optic number of scans:
Bar code quality varies throughout the height of a bar
code. It is best to take multiple scans from top to bottom
and average the scan grades to calculate the overall scan
grade. This parameter can be set from 1 to 10.
The next selection is Auto Optic Days Between
Calibration:
The maximum number of days allowable before the
system must be field calibrated (Auto Optic only).
The system will display a warning message one day
before expiration and after calibration expires inspection
results may be captured but the results will not be stored
in the database and may not be printed.