Press and hold Print button then press Left Arrow button
(at the same time) to display the Calibration screen:
Verifying the Inspector 5000 is calibrated
From the main calibration screen the unit is setup to
measure reflectance. This allows you to check the
calibration of the unit by placing the Auto Optic on the RJS
Calibration Plaque (P/N 002-7410), white or black fields,
and verifying that the reflectance value matches the stated
values.
Verify that reflectance values are /- 2% (of the
value written the calibration plaque
–
“High Reflectance
Value Target Values: Red (660nm)
”)
for white and +/- 2%
(of the value written the calibration plaque
–
“Low
Reflectance Value Target Values: Red (660nm)
”) for
black.