Test duration
Duration between 2 data
points(Sample rate)
80 min 01 sec - 85 min
17 sec/data point
85 min 01 sec - 90 min
18 sec/data point
90 min 01 sec - 95 min
19 sec/data point
95 min 01 sec - 99 min 59 sec
20 sec/data point
Following are the steps to view graph of insulation resistance vs time for a given DUT:
On the home screen go to your desired test mode(IR,DAR,PI,STEP or RAMP) using their
corresponding button.
After this run the test as per the instructions given in the previous chapter.
After the test is completed, press the
button to display the graph window. The
graph window is as shown below:
Cursor
Cursor scrolling
keys
Max.
resistance
during test
Test start time
Test end time
Values on
graph
correspon-
ding to
cursor
position
32