
4.1 EMC Installation Requirements
This product has been designed to meet the certification of the EU directives when
installed to a good code of practice for EMC in industrial environments, e.g.
1. Screened output and low signal input leads or have provision for fitting RF
suppression components, such as ferrite absorbers, line filters etc., in the event
that RF fields cause problems.
NOTE:
It is good practice to install sensitive electronic instruments that are
performing critical functions, in EMC enclosures that protect against electrical
interference which could cause a disturbance in function.
2. Avoid routing leads alongside cables and products that are, or could be, a source
of interference.
3. To protect the product against permanent damage, surge transients must be
limited to 2kV pk. It is good EMC practice to suppress differential surges to 2kV at
the source. The unit has been designed to automatically recover in the event of a
high level of transients. In extreme circumstances it may be necessary to
temporarily disconnect the auxiliary supply for a period of greater than 5 seconds
to restore correct operation.
4. ESD precautions must be taken at all times when handling this product.
18