8.Observe whether water circulation runs well or not. After cooling system runs for at least
5minute, the water has slided into the cooling system and then turn the device off.
If the system sounds beep and show "water current error", then you need to move out the
bubble air from the system. In the stage: switch off the power, take off
the protection plug for water conduit (as Figure 4. 4),
then switch on
the power, meanwhile the water will spout out from the front of water conduit .
Moreover, If a leak is observed, turn the unit off , then take off the protection plug of
water conduit and check whether the O-Ring for faucet is likely to be lost. If the O-Ring
islost , user should take out the O-Ring from accessories and push into water conduit.
9.After cooling system runs for at least 1 minute, the water slide into the cooling system and then
turn off the power.
10. If the water circulation run well. Then switch off the system. It means you can fullyassemble
system with handpiece, screw on the infusion cap and ventilation caps.
11. When installing a new treatment handpiece or switching between treatment handpiece, place
the treatment head lower than the main host in order to remove bubbles from the treatment head,
Then turn the device on. Let the water circulate for at least 1 minute before proceeding with the
treatment.
12. Install multi-treatment head for single head models .
13) Plug the handpiece connector firmly and properly to avoid the water
circulation problem as the following figures.
First : press the gray buttons at the both sides of the handle.
Second : make sure the fixing pins are aiming at corresponding socket.
Third :plug the upper pin directly to the upper socket, then insert the lower one.
as(Figure 4.4)
=
14) Turn the outside circle in the right and fix the handpiece finally as the following figures