User’s Manual

14 of26

8.Observe whether water circulation runs well or not. After cooling system runs for at least
5minute, the water has slided into the cooling system and then turn the device off.

If the system sounds beep and show "water current error", then you need to move out the
bubble air from the system. In the stage: switch off the power, take off
the protection plug for water conduit (as Figure 4. 4),

then switch on

the power, meanwhile the water will spout out from the front of water conduit .
Moreover, If a leak is observed, turn the unit off , then take off the protection plug of
water conduit and check whether the O-Ring for faucet is likely to be lost. If the O-Ring
islost , user should take out the O-Ring from accessories and push into water conduit.

9.After cooling system runs for at least 1 minute, the water slide into the cooling system and then
turn off the power.
10. If the water circulation run well. Then switch off the system. It means you can fullyassemble
system with handpiece, screw on the infusion cap and ventilation caps.
11. When installing a new treatment handpiece or switching between treatment handpiece, place
the treatment head lower than the main host in order to remove bubbles from the treatment head,
Then turn the device on. Let the water circulate for at least 1 minute before proceeding with the
treatment.

12. Install multi-treatment head for single head models .

13) Plug the handpiece connector firmly and properly to avoid the water

circulation problem as the following figures.

First : press the gray buttons at the both sides of the handle.

Second : make sure the fixing pins are aiming at corresponding socket.

Third :plug the upper pin directly to the upper socket, then insert the lower one.

as(Figure 4.4)

=

14) Turn the outside circle in the right and fix the handpiece finally as the following figures

Summary of Contents for QSW500

Page 1: ...Risen Beauty Technology Co ltd Add Room619 Building 1 NO 10 Zhuoxiu North Street Fangshan District Beijing 102401 China User s Manual 1 of 26 Q Switched Nd YAG 1064 532nm Laser Treatment System QSW500...

Page 2: ...afety 10 2 7 Fire Prevention 10 2 8 System Safety 10 2 9 Transport Safety 10 2 10 Safe Marking 11 Chapter 3 Installation 11 3 1 Equipment Detailed List 12 3 2 Installation Requirements 12 3 4 Move 15...

Page 3: ...reet Fangshan District Beijing 102401 China User s Manual 3 of 26 6 4 Purify Soak Lip and Lips Line 25 6 5 Purify Tattoo and Wound Tattoo 25 6 6 Purify Old age Spot and Nevus 25 6 7 Purify Birthmark a...

Page 4: ...procedure Operation Operation of the device Maintenance Maintenance information Clinical Application including application range treatment process and side effect Parameters selection Basis of paramet...

Page 5: ...Device Brief Introduction Model name QSW500 Product name Q Switched Nd YAG Laser System Figure 1 1 shows the device Figure1 1 QSW500 Q Switched Nd YAG Laser System 1 3 Device Structure This device is...

Page 6: ...st shell 4 Pedal control 5 Key switch 6 LCD control panel 7 Plug of Handpiece Front View of Nd YAG Laser Treatment System Notice It is recommended that only the engineer who is authorized by the manuf...

Page 7: ...epeat frequency 1 10HZ Lead light method directly output laser The light spot diameter 2 5mm Power supply 90 130V 50Hz 60Hz or 200 260v 50Hz Environment temperature 5 40 Relative humidity 80 Cooling s...

Page 8: ...y when not in use so as to prevent unauthorized access to the device 2 2 2 Safety Warning Only personnel trained by the distributor are authorized to maintain the inner workings of this device Any man...

Page 9: ...he laser Note Gauze eye shield and tarsi are all effective in the protecting for the eyes 2 5 1 Laser Light Warning The Nd YAG laser is invisible infrared light Among them The laser of 532nm wavelengt...

Page 10: ...stible material such as acetone or alcohol near the device If alcohol based products are used to disinfect the laser device ensure that the alcohol has completely evaporated prior to operation Flammab...

Page 11: ...tion method Key switch The key switch is used shut off power This key is only provided by the manufacturer Please pull out the key after operation Please confirm that the system has been locked Emerg...

Page 12: ...ith adequate ventilation away from devices that produce heat i e refrigerator 3 2 2 Electrical Requirements Before the Laser leaves the factory it has already been marked the local rated voltage accor...

Page 13: ...s bag and unscrew the ventilation screw caps According to the following figure Before every treatment please ensure that there is enough water inside of the machine 4 Ensure that the key is in the off...

Page 14: ...ater slide into the cooling system and then turn off the power 10 If the water circulation run well Then switch off the system It means you can fullyassemble system with handpiece screw on the infusio...

Page 15: ...r the first time rinse the cooling system with distilled water 3 Ensure that the water level is adequate before treatment to prevent over heating 4 Distilled water is added periodically specially when...

Page 16: ...the instrument and handle with care which the instrument should be supported by natural damage The markings such as Handle with care No wetness up should be on the box 3 5 2 Environment 1 Environment...

Page 17: ...in the treatment room must wear the protective filtered glasses supplied 10 Make sure the patient is wearing the opaque protective goggles prior to treatment 11 Avoid looking directly into the intense...

Page 18: ...18 of26 2 Treatment interface figure 5 2 Laser Energy 1 2000mJ Working frequency 1 10 Hz Return to the first interface Start to work Standby Total shooting numbers Current shooting numbers Reset the c...

Page 19: ...ers can all be adjusted when there is something wrong with the machine when the seller give the buyer the background setting password 1 Do not point and operate the device elsewhere other than the int...

Page 20: ...ceeds the required protection temperature the water temperature goes high it will show on the screen at the same time the buzzer will give an alarm and the instrument will cease to work When the tempe...

Page 21: ...ity 3 The damage of system caused by fire the electric power system breaking down lightning flood disaster blast 5 1 Clean Equipment Clean the outer part of the equipment weekly with soft wet tissue Y...

Page 22: ...head is leaks If the laser treatment head is damaged please change the laser treatment head or contact the dealer There is a cheep sound in the instrument operation process when pressing preignition T...

Page 23: ...phage will swallow the function and the transportation function continues long tome The stabbed pigment tissue will became gradually shallow in this period time We suggest the treatment interval of 6...

Page 24: ...e 532nm wavelength directly 5 Purify the scarlet red and brown color eyebrow can use 1064nm wavelength first then use the 532nm wavelength again After 6 weeks use the 532nm wavelength to carry on the...

Page 25: ...as same as purify soak eyebrow 3 The bigger area only purify once time then do not purify again as far as possible Avoid the energy gathering to create the skin swelling or the blister 4 The treatmen...

Page 26: ...y 2 Continuously strike the trouble place with the 1064nm wavelength about 15 minutes 3 The treatment time avoids June July August and September every year 4 The yellow brown spot can be cured by this...

Reviews: