User’s Manual
13 of26
Relative humidity should not be over 80%. The working power of this device is about 2 KW. It
would be best that the Laser unit be installed in an air-conditioned room where the relative
humidity and temperature can be maintained at optimal levels.
3.3 Installation
1.The device should be placed in a room with an indoor temperature of 15~30 C; the humidity
should not be higher than 80%. Operating or treatment rooms must be kept clean of all times.
2.Turn on the Infusion/drain tap then put on the tap connector into the infusion tap from
accessories bag and unscrew the ventilation screw caps. (According to the following figure
)
Before every treatment ,please ensure that there is enough water inside of the machine.
4.Ensure that the key is in the off position and that the emergency cut off button is not engaged.
If it is engaged, release it by rotating the button towards the direction of the arrow indicator.
5.Attach the power cord and attach the foot switch cord.
6.Switch the circuit breaker to the "On" position
7.Switch the device on by turning it in a clockwise direction. At this time water circulation and the
cooling system starts automatically.
Note:
Please always open the
vent (except long dilivery to other
areas)
3. Infuse distilled water until the water level reaches 90%
level of
the water observation window.