28
Reading the Displays
Menu screens
The VM-54 has two menu pages which are shown on the sub display and are
numbered 1/2 and 2/2. The Menu key lets you switch between these pages.
To select an item on a menu screen, use the and keys.
To change the setting of an item, use the and keys.
Note
When an option program is installed, the number of
menu pages is different.
Menu screen 1/2
The display contents will change, according to the sensor type setting.
<Sensor>
Menu1/2
Type
:
Charge
Sensitivity :
X :
0.100 pC/(m/s^2)
Y :
0.100 pC/(m/s^2)
Z :
0.100 pC/(m/s^2)
03/03 10:20:30
<Sensor>
Menu1/2
Type
:
PV-83C
03/03 10:20:30
Menu screen 1/2 (Type: PV-83C)
Sub display
Cursor
Menu screen 1/2 (Type: Charge)
Sensitivity when "Type" is set to "Charge"
Unit: pC/(m/s^2)
Setting range: 0.0100 to 0.0999 0.0001 step
0.100 to 0.999
0.001 step
1.00 to 9.99
0.01 step
10.0 to 99.9
0.1 step
When PV-57(A) is used, set the Z channel sensi-
tivity to 5.10 pC/(m/s^2)